110

Configuring the QAL3 monitoring

Call up the home page of the QAL3 monitoring
Configuring the sample component
Configuring the printer output
EL3060 CONTINUOUS GAS ANALYZERS
ATTENTION
The QAL3 data stored on the memory card is lost if the ECT software tools
("configurator") or the TCT are used to change the configuration of the gas
analyzer or if the memory card is installed in another gas analyzer. In these
cases, the QAL3 data must be exported (see page 114) beforehand.
NOTES
The QAL3 monitoring is deactivated ex works for all sample components, i.e.
no QAL3 data is stored.
If a password (see page 102) has been configured in the gas analyzer, this
password must also be input in the configuration of the QAL3 monitoring.
1
Ensure that the gas analyzer is connected to a PC via the Ethernet port.
2
Open the Web browser on the PC, enter the IP address of the gas ana-
lyzer (if necessary, ask the network administrator) and establish the
connection.
The QAL3 monitoring home page of the EasyLine EL3000 Series is
opened.
1
Click "QAL3 Monitoring" on the QAL3 monitoring home page.
2
Click "Components" (sample components) on the "QAL3 Monitoring"
page.
3
Click the sample component to be configured on the "Components"
page.
4
Enter the following data and values in the "Edit Component" page:
"Active" (QAL3 Monitoring of the sample component activated),
"Installation" (installation location)
"Technician",
"Comment",
"sAMS Zero" (S
AMS
"sAMS Span" (S
AMS
5
Exit the page by clicking "Save" or "Cancel".
1
Click "QAL3 Monitoring" on the QAL3 monitoring home page.
2
Click "Settings" in the "QAL3 Monitoring" page.
3
Check and, if required, change the number of lines on the first page
("Number of lines on first page" - recommendation for portrait format:
35) in the "Printing" window, as well as on the following pages ("Number
of lines on all other pages" - recommendation for portrait format: 50).
4
Exit the page by clicking "Save" or "Cancel".
|