86
Magnos206: Changing the range limits
Menu path
Range limits
Setting range
Measuring range with a suppressed zero point ("suppressed measuring range")
Physical measuring range
EL3060 CONTINUOUS GAS ANALYZERS
Setup
Measurement Ranges
The range limits can also be set in the configurator (see page 82).
NOTES
Measuring ranges must not be set within ignition limits.
The calibration of the measuring range should be checked after the range
limits have been changed.
The start value of the measuring range and the end value of the measuring
range must be within the physical measuring range.
The measuring range can be freely set within the limits of the physical
measuring range. It is set ex works to 0 to 25/100 % vol. of O
The smallest measuring range is 0 to 0,5 % vol. of O
The suppression ratio is max. 1:10, e.g. 19 to 21 % vol. of O
tion with pressure sensor (see page 26) is required.
The analyzer has a physical measuring range. The limits of this measuring
range are 0 % vol. of O
and 100 % vol. of O
2
|
OI/EL3060-EN REV. A
.
2
2
.
2
or per order.
2
. Pressure correc-