Carrying Out A Calibration Reset - ABB EL3060 Series Operating Instruction

Continuous gas analyzers
Table of Contents

Advertisement

Carrying out a calibration reset

Menu path
When should a calibration reset be carried out?
What does the calibration reset do?
Status signal
EL3060 CONTINUOUS GAS ANALYZERS
Maintenance
Basic Settings
A calibration reset should only be carried out if the gas analyzer can no long-
er be calibrated by normal means. A possible cause is e.g. that the gas ana-
lyzer has been calibrated with the wrong test gases.
The calibration of the gas analyzer is reset to the ex-works calibration set-
tings by the calibration reset. Furthermore, the offset drift and the amplifi-
cation drift are electronically reset to the ex-works calibration settings.
The status signal "Maintenance mode" (see page 130) is active during the
calibration reset.
NOTE
The calibration reset cannot be carried out while an automatic calibration is
in progress.
|
OI/EL3060-EN REV. A
Calibration Reset
119

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents