Shutting Down And Packing The Gas Analyzer; Shutting Down The Gas Analyzer; Packing The Gas Analyzer - ABB EL3060 Series Operating Instruction

Continuous gas analyzers
Table of Contents

Advertisement

Shutting down and packing the gas analyzer

Shutting down the gas analyzer

Shutting down
Ambient temperature

Packing the gas analyzer

Packing
Ambient temperature
EL3060 CONTINUOUS GAS ANALYZERS
In the case of a temporary shutdown:
1
Shut off the sample gas.
2
Purge the gas lines and gas feed paths in the gas analyzer with dry fresh
air or nitrogen for at least 5 minutes.
3
Switch off the power supply to the gas analyzer.
In the case of a long-term shutdown, carry out the following in addition:
1
Remove the gas lines from the gas analyzer ports. Tightly seal the gas
ports.
2
Disconnect the electrical leads from the gas analyzer.
Ambient temperature during storage and transport: −25 to +65 °C
1
Remove adapters from the gas ports and tightly seal the gas ports.
2
If the original packaging is not available, wrap the gas analyzer in bubble
wrap or corrugated cardboard. When shipping overseas, additionally
shrink-wrap the gas analyzer air-tight in 0.2 mm thick polyethylene film
adding a drying agent (such as silica gel). The amount of drying agent
should be appropriate for the package volume and the expected ship-
ping duration (at least 3 months).
3
Pack the gas analyzer in an adequately sized box lined with
shock-absorbing material (foam or similar). The thickness of the
shock-absorbing material should be adequate for the weight of the gas
analyzer and the mode of dispatch. When shipping overseas, additional-
ly line the box with a double layer of bitumen paper.
4
Mark the box as "Fragile Goods".
CAUTION
When you return the gas analyzer to the service department, e.g. for re-
pair, please state which gases have been supplied to the gas analyzer.
This information is needed so that service personnel can take any requi-
site safety precautions for harmful gases.
Ambient temperature during storage and transport: −25 to +65 °C
|
OI/EL3060-EN REV. A