Automatic calibration: Manual start

Menu path
Wait until the warm-up phase has ended
Start automatic calibration manually
Abort automatic calibration manually
EL3060 CONTINUOUS GAS ANALYZERS
Operation
Calibration
The gas analyzer may not be calibrated until the warm-up phase has been
concluded.
Duration of the warm-up phase:
Uras26: approx. ½ hour without, approx. 2.5 hours with thermostat
Magnos206: approx. 2 hours
Magnos28: approx. 2 to 4 hours
Caldos27: approx. ½ hour
Caldos25: 1 to 4 hours, depending on the measuring range
NOTE
If the automatic calibration is aborted, the analyzer is in an undefined state
(with respect to the calibration). For example, the zero-point calibration may
have been completed and calculated, but the end-point calibration has not
yet been carried out. As a result, the automatic calibration must be restarted
after it has been aborted and allowed to run through to the end.
|