Manual Calibration: Calibration Method - ABB EL3060 Series Operating Instruction

Continuous gas analyzers
Table of Contents

Advertisement

Manual calibration: Calibration method

Menu path
Calibration method
Analyzer
Uras26
Magnos206, Magnos28
Caldos27
Caldos25
EL3060 CONTINUOUS GAS ANALYZERS
Setup
Calibration Data
The calibration method can also be set in the configurator (see page 63).
Calibration method
Zero-point/end-point calibration
Zero-point/end-point calibration
or
substitute gas calibration
or
single-point calibration
Zero-point/end-point calibration
or
single-point calibration
Zero-point/end-point calibration
or
substitute gas calibration
|
OI/EL3060-EN REV. A
Manual Cal. Method
Remark
cannot be altered
If a substitute gas component has been
installed in the Magnos206 (see page 73)
or Magnos28 (see page 75) analyzer, the
substitute gas calibration is set ex works
as the calibration method; otherwise, the
zero-point/end-point calibration is set.
The single-point calibration can be se-
lected as an alternative.
set ex works
with standard gas (see page 77)
set ex works
can be selected if a substitute gas com-
ponent has been set up in the Caldos25
(see page 79) analyzer
67