5.
Ensure the serial numbers of the following H201Ti's components correspond to those
indicated on the Test Certificate and Data Sheet (for an example, see Figure 6-1 on
page 6-4) and on the shipping box:
• Serial number of the H201Ti (located on the back of the heating plate; see item 2 in
Figure 2-2 on page 2-5)
• Serial number of the CPU module (located on the back of the module; see item 4 in
Figure 2-2 on page 2-5)
• Serial number of the I/O module (located under the module; see item 6 in Figure 2-2 on
page 2-5)
• Serial number of the sensor (located under the sensor's connector; see item 3 in
Figure 2-2 on page 2-5)
Note: Each H201Ti is configured for a specific sensor. If several H201Ti's are received,
take the necessary precautions not to swap the sensors and H201Ti's.
6.
Report any discrepancies to the company supervisor and the General Electric Canada
Customer Service (the coordinates can be found at the bottom of page ii).
Note: Never return equipment without first contacting the General Electric Canada
Customer Service.
6.1.4

Storage and Battery Care

If the Hydran 201Ti is not installed immediately, it is strongly recommended to store it in
its original shipping box in a safe place. The plastic cap must be left on the adaptor to
protect its threads and the membrane.
Storage should be at a temperature between 5 and 45 °C (41 and 113 °F) and a non-
condensing, relative humidity between 5 and 95 %; if these limits are to be exceeded, please
contact the General Electric Canada Customer Service (the coordinates can be found at the
bottom of page ii). These restrictions do not apply to transportation.
When storing the H201Ti for more than six months, it is recommended to disconnect the
lithium battery located inside the CPU module (see item 4 in Figure 2-2 on page 2-5); to
disconnect and reconnect the battery, see Section 2.7.2 on page 2-15. The battery has a
shelf life (when unused) exceeding ten years; if the battery remains connected when the
H201Ti is not powered, operation of the real-time clock and data memory reduces its life
to approximately three years.