Probes Environmental Requirements; Time And Manpower Requirements - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

2-2-5 Probes environmental requirements

2-2-5-1 Operation, storage and transport temperatures for probes
CAUTION
Temperature
Humidity
Pressure
CAUTION
CAUTION
NOTE:

2-2-6 Time and manpower requirements

Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
Probes should be operated, stored, or transported within the
parameters outlined below.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6:
Probe Environmental Requirements
Operational
3° - 40°C
38° - 104°F
30 - 80%
non-condensing
700 - 1060hPa
Check the room temperature before you use the probe.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Refer to
Table 2-1 on page 2-3
settlement time.
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
General Ultrasound system requirements
Storage
-5° - 50°C
23° - 122°F
10 - 90%
non-condensing
700 - 1060hPa
to determine the needed
Transport
-5° - 50°C
23° - 122°F
10 - 90%
non-condensing
700 - 1060hPa
2-11