System Setup
3-9-4-5 DICOM SR
3-56
DICOM Structured Reporting (SR) is a standardized for medical
results. Vivid T9/Vivid T8 supports the specialized form for Adult
Echo Ultrasound (TID 5200 Echocardiography Procedure
Report) and Vascular Ultrasound (TID 5100 Vascular Ultrasound
Procedure Report) for M&A results.
With the DICOM SR support, M&A for an exam can be sent at
the end of the exam or when exported from local archive. The
destination can be either a server on the network or a removable
media (DICOM Media), depending on the DICOM dataflow
selected.
TID5200 Echocardiography Procedure Report is sent if the
exam contains M&A from category Cardiac/Pediatric (heart) and
Vascular/Abdominal categories, two SR documents are sent.
TID5200 Echocardiography Procedure Report and TID 5100
Vascular Ultrasound Procedure Report do not support all M&A
results from Vivid T9/Vivid T8 . They are limited to the following:
No unassigned measurement.
Not modifies Simpson or Bullet methods.
Basic derivations (Average, last, Min and Max), no
references between the derived measurements and the
ones they were made from.
Wall Motion Scoring: individual segment scores only
according to 16-segment model, no graded hypokinesis
(only Hypokinesis is used).
DICOM SR must be activated for each DICOM device.
1. Press Config (F2) on the control panel and log on as admin.
2. Select Connectivity-> Dataflow, the dataflow sheet is
displayed.
3. Select the DICOM dataflow to configure in the Dataflow
pull-down menu.
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3