8-3-4 Internal Battery(continued)
8-3-4-4 Removal Procedure
Table 8-2:
Removal Procedure for Internal Battery (For R2.0.3 and above)
No.
1.
Pull out the body front cover with hands.
2.
Unscrew the 2 screws and remove the
bracket.
3.
Unscrew 1 screw with hand, then
remove the battery bracket. Then take
out the battery.