8-4-7 Loading the System Software

WARNING
NOTE:
8-4-7-1 Loading the System Software with USB memory stick
NOTE:
NOTE:
NOTE:
NOTE:
NOTE:
8-22
While the software install procedure is designed to
preserved data, you should save any patient data,
images, system setups and customer presets to CD,
DVD, USB Flash Drive, or USB Hard Disk before doing a
software upgrade.
Before loading the system software, please ensure that the
power can be continuously supplied and there is no risk of
power cut off during loading procedure.
There is one method to load the system software:
Load the system software with USB memory stick.
While it is believed to be unnecessary, it would not hurt to
disconnect the system from the network and remove all
transducer.
Please ensure AC adapter is connected during system upgrade!
Serial number and option key are needed to be input after
loading the complete disk. The serial number about option key
MUST be the same as the ones before installation. Please
record the serial number and option key before installation.
1. Insert the USB memory stick labeled "System & Application
Software" to the system.
Do not load the system software using the USB port
dedicated for printer.
2. Properly turn off the scanner by momentarily pressing the
Power On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
If the system will not shutdown normally, hold down the
Power On/Off Switch until the light turns off.
3. Press Power button to power on the system and press F7
on the control panel immediately.
Vivid T9/Vivid T8 – Basic Service Manual