Electrical Requirements - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Site Preparations

2-2-3 Electrical requirements

2-2-3-1 General requirements
NOTE:
NOTE:
2-2-3-2 Electrical requirements for the Ultrasound system
2-6
GE Healthcare requires a dedicated power and ground for the
proper operation of its Ultrasound equipment. This dedicated
power shall originate at the last distribution panel before the
Ultrasound system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
In the table below, the electrical specifications for the Ultrasound
system includes monitor and on board peripherals.
Table 2-3:
Electrical Specifications for Vivid T9/Vivid T8 system
Voltage
Tolerance
100-240 VAC
±10%
Power Consumption
Max. 400VA
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
Frequency
50/ 60HZ