3-9-4-5 DICOM SR
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
(continued)
4. Select a DICOM storage device in the Selected devices
and then select Properties.
The Properties window for the selected DICOM storage
device is displayed.
Figure 3-43. DICOM Storage Properties Window
5. Check the option Allow SR to enable DICOM SR.
The following additional options are available:
Allow SR Private Data: send the current exam data in
private format. This option is by default unchecked and
should only be used with DICOM storage devices that
can handle private data format.
Signed Doppler Velocities: send signed Doppler
veloities.
No images:no images are sent, only M&A
Use older SR version:when checked a Use older SR
version pull-down menu is displayed. The current exam
data will be sent in the same format as the selected SR
version. Details about format and content of the SR
version can be found in the corresponding user manual
of the selected version.
6. Select OK.