When There's Too Much Leakage Current; Ac/Dc Fails; Chassis Fails - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Care and Maintenance
10-8 When there's too much leakage

10-8-1 AC/DC Fails

10-8-2 Chassis Fails

NOTE:
10-30
Where applicable, check the AC/DC adapter and its cable.
Replace a new one if any portion is defective.
Check the ground on the power cord and plug for continuity.
Ensure the ground is not broken, frayed, or intermittent. Replace
any defective part.
Where applicable, tighten all grounds. Ensure star washers are
under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other
wiring abnormalities. Notify the user or owner to correct any
deviations. As a work around, check the other outlets to see if
they could be used instead.
No outlet tester can detect the condition where the white neutral
wire and the green grounding wire are reversed. If later tests
indicate high leakage currents, this should be suspected as a
possible cause and the outlet wiring should be visually
inspected.
current ...
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid t9

Table of Contents