7-4-5-13 Connected Probes

Element
Active Probe Temperature (Celsius)
Probe Name
Probe ID
Serial Number
Status
7-34
Connected Probes shows probes connected to the Vivid T9/
Vivid T8 . The order on the user interface is top down matching
the left-to-right order on the Vivid T9/Vivid T8 .
The information on Connected Probes is available to all service
class licenses.
To access Connected Probes, navigate to Utility > Service >
Home.
Figure 7-18. Connected Probes
This table shows all the elements available on Connected
Probes with descriptions.
Table 7-9:
For more information,
Connected Probes
DESCRIPTION
When available, temperature of the active probe.
Not all probes report temperature. The most common probe
to report temperature is the TEE probe.
Name of the probe connected to the Vivid T9/Vivid T8 .
Identifier of the probe connected to the Vivid T9/Vivid T8 .
Serial number of the probe connected to the Vivid T9/Vivid
T8 . If the serial number of the probe is not available, then N/
A displays.
Statuses of the probe connected to the Vivid T9/Vivid T8 .
Valid values are:
• Active
• Non Active
see:7-4-5 'Home' on page 7-15