Grounding
Router
2001989-203C
Battery age
Ambient temperature
Site specific UPS usage patterns
Load characteristics
Your actual run times may be different.
UPS description
Power supply UPS 600VA 120 volts
Power supply UPS 600VA 220 volts
Power supply UPS 750VA 120 volts
Power supply UPS 750VA 220 volts
Power supply UPS 2000VA 120 volts
The UPS part number information is in the parts list in this manual.
components on page 7-2.
To ensure safety and for proper operation, the ATS must be properly grounded in
accordance with all applicable IEC and UL specifications, applicable building codes,
and local electrical inspection regulations. For international installations, grounding
must comply with all internal standards agencies within the specific country.
The common ground point for wall receptacles should be connected to the nearest
equipotential ground through a bonded grounding system. If a bonded grounding
system is not available, the ground pin of each wall receptacle must be individually
connected (using 10 AWG stranded copper grounding cable) to a central grounding
point. Do not jumper the ground pins of the receptacles together.
The grounding system must not carry current, such as a grounded neutral, since the
current flow could produce potential differences along the ground path. Potential
differences are a shock hazard source for equipment users and patients. All grounded
objects in the installation area must be at the same ground potential.
A router allows connection between the hospital network and the Unity Network IX
network. The router is used mainly for remote service connections. For further details
refer to the Unity Router Configuration Document.
Routers may not be required for all ATS installations. Refer to the site survey.
ApexPro™ Telemetry
Runtime estimates
1 ATS for approximately 40 minutes
2 ATSs for approximately 15 minutes
1 ATS for approximately 42 minutes
2 ATSs for approximately 14 minutes
1 ATS for approximately 48 minutes
3 ATSs for approximately 11 minutes
1 ATS for approximately 41 minutes
3 ATSs for approximately 10 minutes
1 ATS for approximately 1 hour, 22 minutes
5 ATSs for approximately 9 minutes
See Optional