SNMP boards

PN 2021512-009
PN 2021512-017/
2021512-021
2001989-203C
CAUTION
LOSS OF CONNECTIVITY — SNMP board replacement requires
re-entry of serial number information. Improper ATS serial number
information will prevent remote services connection.
NOTE
The SNMP battery is not required and should be removed if present. No
functionality will be lost by removing the battery.
If you remove the SNMP battery, you need to do the following after a reboot:
Ensure that the Remote Service configuration is still set properly.
Configure the server for Remote Service access on page 5-10.
Verify that the serial number is correct. From Webmin, select Information >
System Information > System Information.
NOTE
The SNMP board FRU (pn 2021512-017) has been replaced with SNMP board
(pn 2021512-021). Using the pictures below, identify which board(s) are being
used.
1.
Complete the preparation procedures.
7-3.
2.
Remove the chassis cover.
3.
Disconnect the CD data cable connector from the left edge of the CPU module.
Move the CD data cable to out of the way to allow free access to the SNMP
boards.
4.
Disconnect the power cable connector from the CPU module. Move the power
cable to out of the way to allow free access to the SNMP boards
ApexPro™ Telemetry
Parts lists, drawings and replacement
See Field replacement preparation on page
See Chassis cover on page 7-6.
See
NOTE
Ensure that the
connector
differences are
taken into account
if an old board is
being replaced
with a new board.
258A
7-11