Tec 6 Plus Vaporizer
Calibration
Electrical supplies
9-2
À WARNING
Note:
All specifications are nominal and subject to change without notice.
Check the calibration certificate that is included with your Tec 6 Plus
Vaporizer.
The Tec 6 Plus Vaporizer can only be calibrated at a
Datex-Ohmeda Authorized Service Center.
The Tec 6 Plus Vaporizer is calibrated at 760 torr and at a nominal
21 °C for the following concentrations (v/v percent): 1, 2, 3, 4, 5, 6, 7,
8, 9, 10, 12, 14, 16 and 18, using O, as the carrier gas flowing at five
liters/minute.
Voltage requirement
100 to 120V AC 50/60 Hz or
220 to 240V AC 50/60 Hz
Power consumption
250 VA nominal
Current input
Maximum current requirement is 2.0 A
The ~ symbol on the vaporizer rating label denotes that the external
AC power supply must be single phase sinusoidal alternating current.