Installing the battery
and the power cord
Step 1
Remove the battery cover.
*
Invert the vaporizer.
*
Unscrew the battery cover
securing screw.
e
Remove the battery cover.
4-2
The battery terminals must be firmly clipped onto the
battery to help prevent a possible disconnection when the
vaporizer is moved.
Use only a Datex-Ohmeda power cord to connect the
vaporizer to the electrical supply.
When routing the power cord to the electrical supply,
ensure that it does not interfere with the correct functioning
of other equipment.
Do not support the vaporizer by holding the control dial
and/or locking lever. Hold the main body of the vaporizer
with both hands and keep the vaporizer in an upright
position. Failure to do so may cause the vaporizer to
malfunction.
Install only a Duracell MN1604 or VARTA Energy 2000
nine volt battery in the vaporizer. If any other battery is
installed, it can damage the vaporizer.
A new battery and the power cord supplied with the vaporizer must be
fitted during initial installation if they are not already installed.