Step 3
Test the vaporizer output
alarm.
+
Turn the dial to a setting of
1% or above.
e
Unplug the power cord
from its outlet socket.
+
Wait for at least 15
seconds and then check
that both the auditory
alarm and the red NO
OUTPUT light are
activated.
e
if the alarm and light are
not activated, do not use
the vaporizer.
1107-0641-000
1/23/01
15 seconds
|
_ DESFLURSNE
Operating Instructions
Figure 5-8 « Testing the output alarm
5-7