Operational Io; Agent Level Disdlay; Agentleveldisplay - GE Tec 6 Plus Operation And Maintenance Manual

Table of Contents

Advertisement

Tec 6 Plus Vaporizer
Operational
Agent level display
3-8
The green OPERATIONAL light is continuously illuminated when the
vaporizer is in operational mode. The dial can be turned from © when
the vaporizer is operational.
DESFLURANE
MZ
OPERATIONAL
RO DUYANI
Figure 3-9 * Operational
The LCD agent level display consists of a series of bars that indicate
the nominal volume of agent in the vaporizer sump.
The IE symbol means there is approximately 390 ml of agent in the
sump. The |} symbol means there is 60 ml or more of agent in the
sump. No bar displayed means there is 60 ml or less of agent in the
sump.
As the sump fills, the first agent level indicating bar appears in the LCD
volume display when there is between 60 and 80 ml of agent in the
sump. As the level increases, more bars are displayed, as illustrated
in Figure 3-9, and as the level decreases, fewer bars are displayed.
When the bars are either level with or below the < symbol, the sump
can accommodate 240 mi of agent, which is one full bottle. When there
are no bars visible, as illustrated in Figure 3-10, the LOW AGENT level
alarm is activated (see Low agent level).
Note:
The agent level display is calibrated to be most accurate when the
vaporizer is in a perfectly level position. Deviations from a level po-
sition affect the accuracy of the agent level display and may acti-
vate the LOW AGENT level alarm.