Tec 6 Plus Vaporizer
Step 4
If there is sufficient agent in the
DESFLURANE
sump, check that each light is
extinguished, except the
amber WARM-UP
light which
indicates that the vaporizer is
in its warm-up condition. If the
vaporizer is warm before it is
Puy
connected to the electrical
supply, the green
wy
BENT
OPERATIONAL light may
DOL
«
illuminate immediately.
WARM-UP
A
in is
ao
Figure 5-3 « Checking Warm-up light
Step 5
After a warm-up period of up to
DESFLURANE
10 minutes check that the
amber WARM-UP light
AZ
extinguishes and the green
OPERATIONAL
OPERATIONAL
light
illuminates to indicate that the
vaporizer is ready for use.
Approx.
>
if the OPERATIONAL light
ten
minutes
does not illuminate within 10
minutes, refer to the
Troubleshooting guide in
Chapter 8.
Figure 5-4 « Checking Operational light
Step 6
The alarm and indicators can
> 3 sec = Alarm test
be tested at any time by
pressing UW for at least three
N
seconds to activate the alarm
and display test. Activating the
Figure 5-5 « Testing alarm
test does not affect the
operation of the vaporizer.