Tec 6 Pius Vaporizer
Checking the vaporizer for correct mounting
À WARNING
To help to ensure correct operation, do not use a vaporizer
that is either visibly out of line on the manifold or that can
be lifted off the manifold when the locking lever is in the
locked position.
If more than one vaporizer is fitted, visually check to make sure that
the tops of the vaporizers are horizontal. If the vaporizer is visibly out
of line, perform steps 3 and 4 as described in Removing the vaporizer
from a manifold and remount it correctly.
When the vaporizer appears to be level and the locking lever is in the
locked position, attempt to lift the vaporizer straight up from the mani-
fold. If the vaporizer can be lifted off the manifold, it is not correctly
mounted. Remount the vaporizer (see Vaporizer mounting Proce-
dure).