Step 1
Press the dial release and turn
the dial in a counter-clockwise
direction from the © setting.
DIAL RELEASE
Step 2
Release the dial release and
turn the dial to the required
percentage setting.
Figure 5-12 + Dialing the required percentage
Step 3
Check that when the vaporizer
dial is turned from ©, no other
vaporizer mounted on the
same manifold can be tured
on.
1107-0641-000
1/23/01
5-11