Cleaning; External Cleaning - GE Tec 6 Plus Operation And Maintenance Manual

Table of Contents

Advertisement

Tec 6 Plus Vaporizer
User Maintenance
Maintenance
Intervals
Cleaning
À WARNING
External cleaning
À WARNING
6-2
Prior to performing any maintenance procedures or returning to a
service center for repairs, clean and disinfect the vaporizer.
Annually:
Battery replacement.
Two years from
Planned safety inspections together with the
purchase date and
anesthesia system by qualified personnel.
every six months
.
thereafter:
eps
and perform Output concentration
Do not put any substance other than Suprane™
(desflurane) into the vaporizer sump.
Do not immerse the vaporizer in water or any other liquid.
Do not autoclave the vaporizer.
Do not allow cleaning agents to accumulate in the filler, the
gas inlet and outlet ports or around the control dial.
To clean external surfaces, use a moist cloth and a detergent
(pH 7 to 10.5).
For disinfection purposes, cleaning agents of a surface germicide type
should be used. In terms of material compatibility, cleaning agents
.
which contain the following active ingredients are suitable for use:
*
Aldehydes,
e
Alcohol,
*
Quaternary ammonium compounds.
To avoid possible damage to the vaporizer, do not use the
following compounds as cleaning agents:
Halogen hydrolyzing compounds,
+
Strong organic acids,
Oxygen hydrolyzing compounds.
3/23/01
1107-0641-000