12.6 Identifying the Sensor Units

1)
Purpose
In order for the control PC to identify the sensor units connected, the sensor serial number
of each sensor unit is input to the Control PC.
2)
Notes
2-1) These operations must always be implemented at the installation stage and when any
of the Imaging Units (sensor) or Control PC (hard disk) has been replaced or when the
combination of equipment has been changed.
2-2) The sensor serial numbers must always be input. If the serial numbers of the sensor
unit and Control PC do not match, "Sensor Unit: Detect Error (-5100)" will be
displayed on starting up the system. These numbers are the same as what is input to
the
. (Refer to "Checking the sensor serial numbers".)
PWB-Di
3)
Preparations (What to have ready)
Tool keyboard, tool mouse
4)
Procedure
4-1) Start up the CXDI unit.
4-2) Once the normal sensor screen has appeared on the operation unit, use the keyboard to
enter the debugging mode (Use [ALT] + [TAB].).
4-3) "Welcome to CCR" appears. Select "1 Set-Up..."
4-4) "Setting Mode (0:Normal, 1:Expert)[0=0x0]:" appears. Select "0:Normal."
4-5) "CCR SETUP MENU" appears. Select "7 Scan Sensor Setup."
4-6) The "Capture Device Configuration Table" now appears. Input the number of sensor
units connected to the control PC in the underlined (_) section of "Max Capture
Devices" below.
4-7) Enter the serial number to "A/D Board Serial Number for SensorID#1".
@@@@@@@ Capture Device Configuration Table @@@@@@@
Max Capture Devices [1=0x1]:1
@@@@@@@ Capture Device Configuration No.0 (SensorID#1 OPU)@@@@@@@
-------A/D Board Serial Number 0-0 –> 40GC: 0x
A/D Board Serial Number for SensorID#1 [0x15D00001=365953025]:
Custom Type [0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE
5:CASSETTE 14X17] [0=0x0]:0
Field of View Rotation (0:No 1:Yes) [0=0x0]:0
Constant for Exposure Index
---- Need to re-start program to validate this change.
4-8) When "CCR SETUP MENU" appears, press the [Esc] key to return to "Welcome to
CCR."
4-9) Select the command "8 - Exit" from "Welcome to CCR" menu to exit the CXDI
application.
4-10) After "Windows XP desktop" screen appears, start the CXDI application again.
*When restarting the CXDI application, the alert system information management error
"A/D board management information was updated" is displayed, and so click [OK].