Canon CXDI-40G COMPACT Service Manual page 30

Table of Contents

Advertisement

No.
Step
Set the Fixed ROI areas
17
Set exposure parameter table
18
Set annotation
19
Connect the network and set
20
the output destination
Startup settings
21
Check the linearity of the
transferred image density
22
Correct operation unit gamma
23
Install the grid
24
Self-diagnostic testing
25
Radiographic testing
26
Body parts settings
27
Check and perform the system
28
settings
Perform
the
adjustment
and
29
unneeded data
Clean
30
Explain operation to the user
31
Adjust final parameters
32
Backing up valuable data
33
2. Installation
Conditions and checkpoints
- If necessary, to set the ROI area.
- Set it in consultation with the technician.
- Set it in consultation with the technician.
- Install in the Bucky unit, stand or table.
- Check the grid installation angle.
- Use as a fixed grid.
- No errors must be displayed.
- Check the image quality and transfer
operation to printer and storage.
- The engineer in charge must be consulted
prior to performing these settings.
- The engineer in charge must be consulted
prior to performing these settings.
overall
- Check according to the check sheet.
delete
- Delete unneeded data.
- The engineer in charge must be consulted
prior to narrowing down the adjustments to
the final values.
- Back up when setting, and back up the
setting data to FD.
- 10 -
Reference
"Set the Fixed
ROI areas"
"Table Setup
Setting"
"Performing the
Annotation
Setting"
"Network
Connections" in
"Setting"
"Set Up Startup
Menu" in
"Setting"
"Linearity
Check of
Transfer Image
Density" in
"Setting"
"Operation Unit
Gamma
Correction" in
"Setting"
"Image Quality
Check"
Operation
Manual
"Setting"
"Post
Installation
Check" in
"Setting"
Operation
Manual
Operation
Manual
"Backing
Up
when Setting"
and
"Backing
Up Setting Data
to
FD"
in
"Setting"

Advertisement

Table of Contents
loading

Table of Contents