(4) Temperature detection
The ambient temperature is detected by thermal sensors on the PWB, and if the prescribed
temperature has been reached, a warning, etc. is displayed.
A warning appears when the thermal sensor detects 48 °C (exposure remains enabled), and
an error occurs when it detects a temperature of 49 °C or more (exposure is disabled).
(5) Serial communication
Communication with an externally connected general-purpose terminal is enabled under the
RS-232C standards. (Debugging applications)
(6) Ready lamp control
(7) Log recording
The power-on time, number of exposures, exposure sequence and other data are recorded in
the flash ROM. Logs can be read out by serial communication via Ethernet.
(8) Sensor information storage
The data unique to the sensor is stored in the flash ROM. It can be read out to an external
device using commands via Ethernet.
(9) Remote updating
The firmware and HUB41 PLD codes are updated from the control PC via Ethernet.
(10) Voltage levels inside PWB
1
2
3
4
5
6
7
8
9

1.4.4 PCB-40EAD board

The data (14 bits/pixel, 2688 x 2688 pixels) resulting from the A/D conversion of the
LANMIT output signals performed on the PCB-40EAD boards are transferred to the
PWB-Di board.
3. Function
GND side
J011_2
J011_4
J011_6
J011_8
J011_10
C3 + connector
IC14_10
C4 + connector
IC4_10
C5 + connector
IC17_10
C282 + connector
I24_1
+ side
8 [V] or more
J011_1
5.15-6.5 [V]
J011_3
8 [V] or more
J011_5
20 [V] or more
J011_7
–10 [V] or less
J011_9
3.3 [V] ± 2 %
1.8 [V] ± 2 %
1.5 [V] ± 2 %
5 [V] ± 2 %
- 5 -
Rating

Advertisement