1. Caution

1.1 Caution during Operating

Please pay attention to the following points when installing the equipment.
(1) If the equipment is hoisted, lowered or transported, it must be supported at both sides by a
minimum of two people so there is no danger of it falling.
(2) When transporting the equipment with a forklift, lower the load to the lowest possible
position that does not impede the running of the forklift.
(3) When installing the equipment, be sure the site meets the following criteria:
1) There must be no dripping water in the area.
2) The environment must be free of harmful elements, such as humid or acidic air, air with a
saline or sulfur content, where there is poor ventilation, or where air pressure or temperature is
abnormal.
3) The equipment must not be placed at an angle or subjected to vibration or shock (this
includes during transportation).
4) The equipment must not be kept where chemical products are stored or where gasses are
generated.
5) The site's power supply must be of the correct voltage and frequency for the equipment.
6) The site must be connected to a fully earthed cable with sufficient ground resistance to
meet standard values.
(4) After installation, be sure to dispose of waste product packaging with care and with full
respect for the environment.
(5) As the imaging unit is easily portable, take special care that it is not knocked, dropped or
subjected to strong shocks.

1.2 Installation Restrictions

(1) Install so that the distance between the imaging unit and power box is at least 300 mm from
the imaging unit sides (in four directions) and at least 150 mm from locations other than
the imaging unit sides.
(2) Except for the sensor cable connected to the imaging unit and the cable for the ready lamp
unit, the cables must not be used with moving parts.
(3) Before starting the installation work, discharge any static charge that has built up on your
body. Also, be sure that the static charge has been discharged before touching (removing)
the PCBs and cable connectors.
(4) Install the imaging unit, grid, and phototimer to a Bucky unit, stand or table. The parts that
secure the imaging unit, grid, and phototimer are obtained from the distributor.
(5) Do not remove the imaging unit from the Bucky unit, stand or table for cassette-type
portable usage.