Connecting; Electrical Connections; Safety Information For Rack Device; Rack-Mounted Device - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting

5.1

Electrical connections

5.1.1

Safety information for rack device

5.1.1.1

Rack-mounted device

Note
If auxiliary power is required, check that it corresponds with that on the nameplate and with
the test certification valid for your country.
5.1.2

Safety information for wall-mounted device

5.1.2.1

Danger due to electrical current

Danger due to electrical current
Check whether the existing line voltage corresponds to that specified on the nameplate of
the device, to the technical specifications and to the test certificate valid for your country.
5.1.2.2
Danger due to electrical current
Danger due to electrical current
The power cable to be connected at the customer location could be damaged by incorrect
fitting in the device.
You should therefore carry out a visual inspection and PE conductor test of the cable in
accordance with the country-specific safety standards, regulations or other applicable
directives before starting up the device. Because part of the cable is in the device, always
adhere to the product safety standard IEC/EN 61010-1 or its national equivalent.
Quick Start
Compact Operating Instructions, 05/2015, A5E31805656-03
WARNING
WARNING
5
43