Combination operation
3.3 Gas connections
OXYMAT 7
Table 3-1
Operating mode
Pressure operation
Suction operation
Combination operation with ULTRAMAT 7
Table 3-2
Operating mode
Pressure operation
Suction operation
30
Mounting conditions for the sample gas restrictor: OXYMAT 7
Sample gas restrictor/clamping screw
Operation with one
OXYMAT 7
In front of the sample
gas inlet
In the sample gas outlet In the sample gas out‐
Mounting conditions for the sample gas restrictor: ULTRAMAT 7
Sample gas restrictor/clamping screw
Combination operation OXYMAT/UL‐
TRAMAT
● OXYMAT 7 must always be
purged first.
● Install sample gas restrictor in
front of the sample gas inlet of the
OXYMAT 7. Install clamping
screw on reference gas inlet/gas
outlet.
● Install sample gas restrictor in the
sample gas outlet of the
OXYMAT 7. Install clamping
screw on reference gas inlet and
sample gas inlet.
Operation with two
OXYMAT 7 in series
In front of the sample
gas inlet into which the
sample gas is intro‐
duced first. Replace the
sample gas restrictor at
the inlet of the rear, sec‐
ond module with a
clamping screw.
let, from which the sam‐
ple gas is first sucked
out. Replace the sam‐
ple gas restrictor at the
inlet of the rear, second
module with a clamping
screw.
Combination operation ULTRAMAT/
ULTRAMAT
● No measures.
● No measures.
Compact Operating Instructions, 05/2015, A5E31805656-03
Operation with two
OXYMAT 7 in parallel
In front of each sample
gas inlet
At each sample gas out‐
let
Quick Start