Siemens SIPROCESS GA700 Quick Start Manual page 35

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 3-5
Case
PM
Un‐
change
d
Un‐
change
d
Un‐
change
d
Un‐
change
d
Un‐
change
d
Table 3-6
Case
PM
New
New
Serial number stored in the PU is also the serial number of the device
1)
Quick Start
Compact Operating Instructions, 05/2015, A5E31805656-03
Plug & Measure - Installation/replacement of an AM
Description
AMx
Re‐
The AM is to be replaced by a module
placed
of the same type. The replacement AM
has a different serial number than the
AM connected previously.
Replacement of the AM electronics, on‐
ly permitted without own serial number
stored in the AM electronics
Re‐
AM electronics installed as "dedicated
placed
spare part". Service event.
Re‐
An AM is replaced by an AM of a differ‐
placed
ent type.
New
An additional AM of the same type is to
be installed.
An additional AM of a previously instal‐
led type is installed again.
New
An additional AM of a type that has not
been connected before is to be installed.
Plug & Measure - Installation/replacement of a PM
Description
AMx
Un‐
The PM electronics is replaced. Quota‐
change
tion marks are displayed instead of the
d
serial numbers. At least one AM re‐
mains in the device.
Un‐
PM electronics with new serial number
change
is installed.
d
Combination operation
3.4 Plug & Measure: Permissible activities
Data synchronization/Action
User action. Confirm <ENTER> or re‐
ject <ESC>.
The data is transmitted after confirma‐
tion, except the identification data and
the calibration parameters.
User action. Always confirm <ENTER>.
The data is transmitted after confirma‐
tion, including identification data and
calibration parameters.
Normal startup without message. The
default data of the AM set in the factory
is applied.
Automatic data synchronization, no ac‐
tion required by user.
User action. Confirm <ENTER> or re‐
ject <ESC>.
The data is transmitted after confirma‐
tion, except the identification data and
the calibration parameters.
Automatic data synchronization, no ac‐
tion required by user.
Data synchronization/Action
Automatic data synchronization, no ac‐
tion required by user.
When there are two AMs in the device,
the new PM always synchronizes with
AM1.
The PM receives the data of the previ‐
ously used PM, including its serial num‐
ber
1)
Automatic data synchronization, no ac‐
tion required by user.
In this scenario, the replacement of the
PM electronics results in a loss of data
of the previously installed PM. In addi‐
tion, the serial number is overwritten.
35

Advertisement

Table of Contents
loading

Table of Contents