General Ultrasound System Requirements; Contents In This Section; Ultrasound System Environmental Requirements - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

2-2 General Ultrasound system

2-2-1 Contents in this section

2-2-2 Ultrasound system environmental requirements

2-2-2-1 If the Ultrasound system is very cold or hot
CAUTION
Degree C
-4.5
Degree F
23.9
hours
3
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3

General Ultrasound system requirements

2-2-2 'Ultrasound system environmental requirements' on
page 2-3
2-2-3 'Electrical requirements' on page 2-6
2-2-4 'EMI limitations' on page 2-9
2-2-5 'Probes environmental requirements' on page 2-11
2-2-6 'Time and manpower requirements' on page 2-11
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 2-1:
System Acclimation Time Chart
-2
0.5
3
28.4
32.9
37.4
2
1
0
requirements
40
42.5
45
104
108.5
113
0
1
2
47.5
50
55
117.5
122
131
3
4
6
60
140
8
2-3

Advertisement