Connectivity Setup; Tcp/Ip Screen - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

3-9-1 TCP/IP Screen

Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3

3-9 Connectivity setup

1. Press F2(Config) on the control panel and login as admin,
refer to
4-2-5 'Logging on to Vivid T9/Vivid T8 as "ADM"' on
page
4-10.
2. Select Connectivity on the screen.
3. Select TCP/IP tab, the screen gives an overview of the
network settings for Vivid T9/Vivid T8.
Figure 3-28. TCP/IP Screen
Computer Name: For Vivid T9/Vivid T8 , it is of this form:
VIVIDX-00NNNN, NNNN is the system's serial number.
IP Address.
Server Config: List of servers and buttons to add,
modify or remove servers.
Connectivity setup