Diagnostics/Troubleshooting
7-5-2 Service Diagnostics(continued)
CAUTION
7-5-2-11-1 Probe Pass/Fail Criteria
7-46
Reboot the system after the probe assessment.
Pass
No dead element
Dead elements on the edge (#0, #1 or #end-1, #end)
Fail
Max 2 weak elemets on the scan line: -3dB<elements<-2dB
Any one of the elements in one scan line: element<-3dB
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3