3-9-2-2 To Add a Server in the Server Config List
3-9-2-3 To Modify the Setup for a Server in the List
NOTE:
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
2. Add the server's name in the Server Name field.
3. Add the server's IP address in the IP-Address field.
4. Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right for the Check
button indicates the result:
A green check mark next to the Check button indicates that
the IP-Address is found on the network.
A red check mark indicates that the IP-Address can't be
found.
Possible reasons:
No network connection between the ultrasound system and
the server.
The wrong IP address.
Both the pre-defined servers and new servers can be modified.
Follow this procedure to modify the setup for a server in the
list:
1. Highlight the server you want to modify.
2. Select Modify.
3. Update the information as needed.
4. Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right for the
Check button indicates the result:
A green check mark next to the Check button indicates
that the IP-Address is found on the network.
A red check mark indicates that the IP-Address can't be
found.
Possible reasons:
No network connection between the ultrasound system
and the server.
The wrong IP address.
5. Select OK to confirm your new settings, or Cancel to leave
without doing any changes.
Connectivity setup