3-8-3-5 Wireless network adapter
3-8-3-5-1 Tools
3-8-3-5-2 Manpower
3-8-3-5-3 Preparations
3-8-3-5-4 Connect the adapter using the USB-docking
3-8-3-5-5 Connect the adapter directly to the USB-port
NOTE:
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
No special tools needed.
One person 10 min.
1. Unpack the wireless network adapter.
2. Ensure no physical damage.
1. Attach the Velcro tape strips to USB-docking and plastic
cover:
a. Attach one piece of Velcro tape to the underside of the
USB-dock.
b. Attach the other piece on the left or right side of plastic
cover. Ensure that the Velcro tape is placed so it only
covers front plastic cover, tape on plastic cover must
also be placed as close to the right/left edge as
possible, to avoid conflict with upper UI.
2. Insert the wireless network adapter into the USB-dock,
place the dock onto the Velcro strip on plastic cover to
fasten it. Connect docking USB-port to rear I/O USB-port on
the ultrasound system.
3. Ensure that the wireless network adapter will not interfere
with motion of UI by lowering it to bottom position and swivel
to left/right. Reposition USB-dock if needed.
Connect the wireless network adapter directly into rear I/O
USB-port
Do not use any type of wireless adapter other than a
GE-approved adapter.