Hitachi CR22N Instruction Manual page 22

High-speed refrigerated centrifuges
Table of Contents

Advertisement

2-2-1 Setting Run Conditions
• How to set speed, running time, and temperature
Here are some examples and descriptions:
Setting item
Typical setting
1
Press the SPEED,TIME,
or TEMP area to show
the on-screen keypad.
The last entered value
2
is displayed in each
area. Press the area
of the desired item
again if the color of
the
upper
white.Go to procedure
3 if the color of the
upper part is blue.
3
Enter the desired
numeric value with the
on-screen keypad.
Entered numbers are
moved to the left
every time a new
number is entered.
Make a check, then if
4
you want to change
other setting, press
the area of the desired
item. If you do not
want to change any
other setting, press
the
Enter
Use the
button to cancel input.
Setting range and units
RPM
(SPEED)
22,000rpm
part
is
Color:Blue
2
2
The last one digit is
fixed.
Set it to 22,000rpm.
button.
CE
1.Can be set to any value in
the range from 300 rpm to
9,990 rpm in increments of
10 rpm.
2.Can be set to any value in
the range from 10,000 rpm
to maximum speed
in increments of 100 rpm.
Running time
(TIME)
2 minutes 30 seconds
Color:Blue
:/-
2
0
0
Press the
button to turn the
number of minutes
blue.
For a continuous
run, press the
button.
Set it to 2:30.
(2 minutes 30 seconds)
Can be set to any value up to
99 minutes 59 seconds in
increments of 1 minute.
(Option:Can be set to any
value up to 99 hours 59
minutes in increments of 1
minute.)
2-7
Temperature
(TEMP)
4 °C
Color:Blue
3
0
:/-
Hold
Set it to 4
Can be set to any value in
the range from -20 °C to
40 °C in increments of 1 °C.
4
°C