Panda iRes Warmer
M1110737 006
Screen
© 2007 General Electric Company, all rights reserved
Description
Step
3
You will be prompted to check "Level bed..."
to ensure that the bed is level during the
weighing procedure.
If the bed is not level, then level the bed by
sighting the bubble level on the side bedside
panel and using the bed tilt control lever.
If the bed is level, proceed to the the next
step.
4
In order to "zero" the scale, the display will
prompt you to lift the baby by showing the
"Lift Baby" icon while a tone sounds. Lift the
baby and any tubing or leads attached to the
baby. Make sure that arms, legs, blankets,
and clothing are clear of the mattress.
3-19