Alarm Table - GE Panda iRes Warmer Operation And Maintenance Manual

Table of Contents

Advertisement

Alarm Table

Alarm
Check Baby
Check Baby -
Heat off
Baby Cold -
Check Temp.
Probe
Baby Hot -
Check Temp.
Probe
Confirm Probe
Jack Connection
Temp. Probe
Failure
* 1= high priority audio signal
2= low priority audio signal
Panda iRes Warmer
M1110737 006
Activation Criteria
Radiant % power @
12 min.
100% for > 12 min.
Radiant heater %
12 min.
above "check baby
alarm disabled" limit
for > 12 min.
Radiant % power @
15 min.
100% for >15 min. and
the "Check Baby" alarm
has gone unanswered
for three minutes.
Radiant heater %
15 min.
above "check baby
alarm disabled" limit
for > 15 min. and the
"Check Baby" alarm
has gone unanswered
for three minutes.
<2.0° C from Set Temp.
5 min.
<1.0° C (<0.5° C) from
12 min.
Set Temp.
>2.0° C from Set Temp.
5 min.
> 1.0° C (>0.5° C) from
12 min.
Set Temp.
No longer getting a
2 min.
reading from the tem-
perature probe.
Two thermistors in a
2 min.
probe differ by 0.5° C or
more.
© 2007 General Electric Company, all rights reserved
Alarm Silence
Audio
Signal*
2
2
1
1
1
2
1
2
2 to 1 (after
30 secs.)
2 to 1 (after
30 secs.)
Mode
Baby Mode
Manual Mode
Baby Mode
Manual Mode
Baby Mode
Baby Mode
Baby Mode
Baby Mode
Baby Mode or Manual Mode
when Manual Temperature
Alarm set.
Baby Mode or Manual Mode
when Manual Temperature
Alarm set.
1-21