Check Baby-Heat Off
Confirm Probe Jack Connection
Panda iRes Warmer
M1110737 006
Screen
Screen
© 2007 General Electric Company, all rights reserved
Description
Step
1
Baby Mode
A continuous tone sounds after the radiant
heater has been running at 100% power for
more than fifteen minutes, and the "Check
Baby" alarm" has gone unanswered for
three minutes. The radiant heater then shuts
off automatically and the high priority "Check
Baby-Heat Off" alarm activates.
Manual Mode
A continuous tone sounds after the warmer
has been operating at greater than 25%
heater power* for more than fifteen
minutes, and the "Check Baby" alarm has
gone unanswered for three minutes. The
radiant heater then shuts off automatically
and the high priority "Check Baby-Heat Off"
alarm activates.
Action: silence the alarm and check the
attachment of the skin probe to the baby
and monitor the baby's temperature closely.
*Default power setting is 25%, but this can be
adjusted on the Service Screen. Refer to the Service
Manual.
Description
Step
1
Baby Mode
The
"Confirm Probe Jack Connection"
alarm activates when the skin probe jack is
unplugged.
Manual Mode
It will also activate when there is no probe
in the jack in the Manual Mode, if the Man
Temp Alarm has been set.
Action: check that the skin probe is fully
inserted into the temperature probe jack. If
this does not cancel the alarm, check if the
Man Temp Alarm has been set. If the alarm
is still not cancelled, then replace the temp
probe.
1-19