Temperature regulation alarms

Baby Hot - Check Temp Probe
Panda iRes Warmer
M1110737 006
Screen
© 2007 General Electric Company, all rights reserved
Description
Step
1
Baby Mode
The "Baby Hot - Check Temp Probe" alarm
activates when the baby's skin tempera-
ture is greater than 1.0°C above the Set
Temperature. Silencing the alarm suspends
the low priority audio alarm for twelve (12)
minutes. When the baby's temperature is
greater than 2.0°C above the Set Tempera-
ture a high priority alarm activates. Silencing
the alarm suspends the high priority audio
alarm for five (5) minutes.
The low priority alarm setting can be nar-
rowed to 0.5°C using the "Setup" screen and
"Pt Temp Alarm" keys.
Manual Mode
The alarm can be activated in the Manual
Mode using the "Setup" and "Man Temp
Alarm" keys. (See Chapter 3 for detailed
description.)
Action: monitor the baby carefully and
check the attachment of the probe on the
baby's skin.