Siemens SIPROCESS GA700 Compact Operating Instructions page 192

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

Technical specifications
8.2 General technical specifications
Influence variables
Sample gas pressure
Sample gas flow
Accompanying gases
Supply voltage
Table 8-19
Climatic conditions
Storage and transport
Max. permissible ambient temperature surround‐
ing the basic device during operation
Ambient humidity (rel. humidity) during storage,
transportation or operation
Restriction for installing an ULTRAMAT 7 analyzer module: 5 to 45 °C
1)
Avoid condensation on the installed components!
2)
Table 8-20
Gas connections
With pipes
86
Deviation at zero point
Deviation of the span
gas
Deviation at zero point
Deviation of the span
gas
Analyzer module
24 V DC (-5%/+10%)
Basic device 85 V to
264 V AC / 47 Hz to
63 Hz
OXYMAT 7: Climatic conditions
1)
OXYMAT 7: Gas connections
≤ 0.2% of the smallest span / 1% or ≤ 50 vpm
O
/1%, whichever is greater
2
Deviation ≤ 0.2% of current span / 1 % pressure
change or 50 vpm O
/1% pressure change, which‐
2
ever is greater
≤ 1% of smallest span per 0.1 l/min change in flow
or 50 vpm O
per 0.1 l/min change in flow within
2
the permissible flow range, whichever is greater
≤ 1% of current span per 0.1 l/min change in flow
or 50 vpm O
per 0.1 l/min change in flow within
2
the permissible flow range, whichever is greater
Zero point deviation (cross-sensitivity) in accord‐
ance with Table A.1 of EN 61207-3
< 0.1% of the current span
< 0.1% of the current span
- 30 to 70 °C
0 to 50 °C
< 90 %
Connecting socket for pipe with ∅ 6 mm
Devices in explosion-proof design
Compact Operating Instructions, 05/2015, A5E35134047-02