Calibrate The Ph Monitor - GE AKTAavant Getting Started

Hide thumbs Also See for AKTAavant:
Table of Contents

Advertisement

7 Prepare the system for a run

7.5 Calibrate the pH monitor

7.5 Calibrate the pH monitor
Introduction
If pH will be measured during the chromatographic run, the pH monitor should be
calibrated before the run is started. Use two pH calibration buffers with a difference of
at least one pH unit. Preferably use a pH standard buffer pH 4 or 7 as the first calibration
point, and a pH standard buffer close to the lowest or highest pH you need to measure
as your second point. Allow the buffers to reach the operating temperature before use.
Note:
Calibrate the pH monitor
Follow the instruction below to perform the calibration.
Step
1
2
3
4
88
Do not run a system flow during pH calibration.
Action
Open the System Control module and select System:Calibrate.
Result: The Calibration dialog opens.
Set the pH monitor as the Monitor to calibrate by selecting pH from the list.
Click
Result: The pH valve switches to the calibration position.
Enter the pH of the first pH standard buffer in the pH for buffer 1 field.
.
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB