Step
3
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
Action
In the Peak Integrate dialog:
Select which curve to integrate (e.g., UV1_280 for proteins).
•
Select a target peak table for storage of the result.
•
Select Calculate baseline from the Baseline list.
•
Click OK.
•
Result: The Peak Table is displayed beneath the active chromatogram, and
the start and end points of each peak are marked by vertical marks in the
chromatogram.
9 Evaluate and print the results
9.2 Peak integration
121