Figure 4-5 Typical Chromatograph For Chrom-1 (Heavies) - ABB NGC8206 User Manual

Chromatograph
Hide thumbs Also See for NGC8206:
Table of Contents

Advertisement

2) When the current cycle completes, the unit should begin a calibration on the
designated cal stream (stream 4 default).
3) When the calibration is complete, the unit should move to the designated next
mode. View results on the Analyzer Operation screen.
4) The calibration stream's Un-Normalized total should be 6.5% (between 99.5%
and 100.5%).
5) Carefully examine the calibration stream's Chrom-1 and Chrom-2 by clicking the
and Figure 4-6 should be used as a guide.
6) Once the unit is running smoothly and is producing good Chroms and all the
peaks are labeled and eluting correctly, perform a save and restore procedure to
update TFCold.

Figure 4-5 Typical Chromatograph for Chrom-1 (Heavies)

2101510 Rev. AG
If values exceed these parameters, proceed to the
Troubleshooting
button. Look for unlabeled peaks and base line anomalies.
If errors exist, proceed to the
manual.
section of this manual.
Troubleshooting
Figure 4-5
section of this
Page 4–11

Advertisement

Table of Contents
loading

Table of Contents