GE ApexPro TT Technical Manual page 78

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Maintenance
4-16
4.
Configure the ASCII setup.
a.
Select File > Properties.
b.
In the Settings tab, click ASCII Setup...
c.
Verify the settings are the same as shown:
d.
Click OK.
5.
Press Enter.
6.
When prompted, type the password mms_aps and press Enter.
7.
At the @> prompt, type ss and press Enter.
8.
Determine if patients are admitted to the receiver, use the following graphics for
reference:
If the column next to FQ is all -1 values, no beds are admitted on the receiver
subsystem tested. Proceed to step 9.
ApexPro™
Baud-Rate
19200
Data Bits
8
Parity
None
Stop Bit
1
Flow Control
Xon Xoff
401A
2001989-351A