Device Specifications; Bias Tee Specifications; Environmental Specifications - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Storage environment

Device specifications

Bias tee specifications

Environmental specifications

Device specifications
2001989-351A
Temperature
-40 – 75°C (–40°F to 158°F)
Humidity
10 -95% non-condensing
Isolation
Meets IEC 60601, classification BF, UL544 patient care, CSA 125
risk class 2G
Overall regulation
< 5% no minimum load required
Maximum ripple
< 100mVp-p
Cord length
305mm (12 in.)
Safety
Approved to UL 544/2601.1, cUL (CSA) 22.2 #125/601.1, TUV
EN60601.1 and CE LVD
EMC
Meets level B requirements of FCC part 15, CISPR11 (EN55011).
Temperature
0 - 50° C (32°F to 122°F)
Frequency
400-650MHz
Insertion loss
0.5dB, max
Isolation
20dB, min
DC OUT/RF IN port
to DC IN port
RF OUT port to DC
IN port
VSWR
1.3.1, max
DC voltage
30 Volts, max
DC current
1 Amp, max
RF power
+20dBm, max
DC resistance
0.1 ohms