Setup Antenna Fields - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Ethernet
Async
Comm
Power
Switch
AC Power Cord

Setup antenna fields

2001989-351A
If the distance between the ApexPro Telemetry System and the Receiver System
is less than 100 meters (328 ft.) use point-to-point with Ethernet crossover cable
connection.
If the distance is greater than 100 meters (328 ft.) then use either an additional
hub or use fiber optic cable.
1.
Connect the RJ-45 (RX network) to the Ethernet port.
2.
Connect the coaxial antenna cables to the antenna inputs. Unused antenna input
jacks need to be terminated with a 75 Ohm F-type male terminator.
3.
Connect the power cord to the AC power inlet. Secure the cord with the strain
relief. Plug cord into emergency AC power outlet.
4.
Indicate which ApexPro Telemetry System is directly connected by labeling the
Receiver System with the ApexPro Telemetry System name. Place the label near
the RX Ethernet port. If desired, this step may be omitted as the ApexPro
Telemetry System can also be identified by using the Blink Rack command at the
ApexPro Telemetry System.
5.
Switch the power switch to I (on).
Strain Relief
Use the following procedure for configuring receiver subsystem so that it listens to
antenna fields that are setup.
NOTE
The factory default is that all four fields are enabled.
1.
Using the 9-pin, serial cable supplied with the transmitter programming kit,
connect a PC to Async Comm (asynchronous serial communication) for setup.
ApexPro™
Installation and configuration
Equipotential
Ground
Antenna
Connectors
RearPanel
3-15