Replace The Fuse; Open The Unit - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Hardware

Replace the fuse

Open the unit

2001989-351A
Observe the following guidelines when handling all PCB assemblies.
Take precautions against electrostatic discharge damage.
Handle all PCB assemblies by their edges.
Before disassembly, note the positions of any wires or cables, marking them if
necessary to ensure that they are replaced correctly.
Gray ribbon cables have retainer clips holding them in the connector.
Save and set aside all hardware for re-assembly.
1.
Open the door on the AC inlet module to access the fuse holder.
2.
Remove two fuses and replace with new on each side of the holder.
3.
Close the AC inlet module doors.
CAUTION
EQUIPMENT DAMAGE—Power must be off to add or remove any
internal assemblies or circuit boards.
1.
Turn the unit OFF at the rear power switch and disconnect the AC power cord
and all communication cables.
2.
Remove 2 screws from the front cover.
3.
Remove cover and set aside.
ApexPro™
Replaceable parts
6-25